# NATA's 2026 NDT Annex: What Changed for PAUT, TOFD, CR and DR Software

> NATA's NDT annex to ISO/IEC 17025, issued May 2026 and effective August 2026, puts PAUT and TOFD software under validation and version control, asks CR/DR procedures to control filters, and makes Level 3 endorsement for advanced techniques a requirement. Restated in plain English, checked on 13 November 2026.

By Dom O'Carroll. Published 2026-11-16; facts last checked 2026-11-13. Canonical: https://subfractal.com.au/thinking/nata-ndt-annex-2026

## The short version

- NATA's NDT annex to ISO/IEC 17025 (SAC 03.3) was issued in May 2026 and took effect in August 2026.
- For PAUT and TOFD, the annex says software used for data acquisition and analysis must be validated and version controlled, and any update must be traceable and assessed for its impact on inspection performance.
- Display and analysis software is checked as updated or annually: verify the software revision and confirm consistent performance for flaw sizing and reporting.
- CR/DR procedures must include controls on digital processing functions such as filters and contrast adjustments, and each job record must state the processing parameters and the software and version used.
- The person overseeing PAUT, TOFD or CR/DR on welds must hold an independent endorsement for those techniques equivalent to ISO 9712 Level 2 or higher; NATA's amendment table says this was upgraded from an expectation to a requirement.

If you run PAUT, TOFD, CR or DR under NATA accreditation, the rules for your software changed in August 2026. They are in NATA's NDT annex to ISO/IEC 17025, a 48-page PDF. This page restates the parts about software, records and technical control, with the section each comes from.

Everything below is taken from NATA's annex. Where we quote, we quote NATA. Read the annex itself, and ask your NATA client coordinator where anything is unclear.

## What is the annex, and when did it take effect?

The annex sets out NATA's specific criteria for non-destructive testing under ISO/IEC 17025. It sits under NATA's infrastructure and asset integrity criteria and covers radiography, ultrasonics, magnetic particle, penetrant, visual and electromagnetic testing, plus work sites. The 2026 issue was issued in May 2026 and took effect in August 2026.

NATA's amendment table lists editorial changes throughout, plus a handful of specific ones. For anyone running advanced techniques, the ones that matter most are about people and software.

## What does it say about PAUT and TOFD software?

In its requirements for advanced ultrasonic testing equipment (Appendix B), the annex lists what a facility must ensure for PAUT and TOFD. One line covers software:

> Software used for data acquisition and analysis is validated and version controlled. Any updates must be traceable and assessed for impact on inspection performance.

Its equipment-check table then sets a frequency:

| Item | When | What NATA's annex asks |
|---|---|---|
| Display/analysis software | As updated or annually | Verify the software revision. Confirm compatibility and performance consistency for flaw sizing and reporting |

Read together, that is three things:

1. **Know what version you are on**, for acquisition and for analysis.
2. **Treat every update as a change.** Record it, and assess whether it could affect detection, sizing or reporting before you rely on it.
3. **Check at least once a year anyway**, even if nothing changed.

Analysis software gets updated, sometimes automatically. If the update lands mid-project, the question an assessor could reasonably ask is which version produced which result, and how you know the new one sizes the same way.

## What does it say about CR and DR?

For CR and DR techniques, the annex (Appendix A) says procedures must cover image acquisition and processing, including:

> controls around the use of digital processing functions (e.g. filters, contrast adjustments)

and requirements for validating techniques, "especially where they are not based on a published standard". The same list asks for documented exposure-parameter selection and image quality requirements: minimum SNRN, sensitivity and basic spatial resolution.

A filter is a judgement made by software. If an operator can sharpen, smooth or stretch contrast before evaluating an image, the procedure should say which settings are allowed, and the record should say which were used.

## Which job records now name the software?

Each method's appendix lists the minimum fields a job record must include "to ensure that the test can be replicated, if necessary". The software-related fields are below. Each appendix has other fields too.

| Method | Software-related fields NATA's annex lists |
|---|---|
| Radiography, CR/DR (App. A) | Detector type/model, pixel size, BSR, minimum SNRN achieved, image processing parameters/filters used, software and version used for acquisition and analysis |
| Ultrasonics, PAUT/TOFD (App. B) | Scan plans and coverage, scan type, focal law parameters, scanner/encoder parameters, software version and analysis method/settings |
| Electromagnetic testing (App. F) | Equipment settings including filters and mixing, detailed instrument setup files and software configuration parameters, software version and settings used |

If your report template or job sheet has no field for software version, that is the simplest fix in the whole annex.

## What changed for the Level 3?

Two things.

**Endorsement for advanced techniques is now a requirement.** Where the work includes PAUT, TOFD, or CR and DR for welds, the annex says the person responsible for overseeing those tests "shall hold an independent endorsement for these techniques that is equivalent to ISO 9712 Level 2 or higher". NATA's amendment table says this was "upgraded from an expectation to a requirement".

**Level 3 coverage is explicit.** The annex says the person responsible for technical control must have direct and ongoing involvement in eight activities:

| # | Technical-control activity | Can be delegated to Level 2? |
|---|---|---|
| i | Authorising technicians to perform tests and release results | No |
| ii | Writing, approving or reviewing procedures, and validating them | No |
| iii | Monitoring the need for new procedures | No |
| iv | Managing the in-house NDT competency program | No |
| v | Monitoring the quality of reporting across the organisation | Yes |
| vi | Ensuring field technical control | Yes |
| vii | Dealing with technical queries from testing personnel | Yes |
| viii | Ensuring the integrity of equipment commissioning and performance checks | Yes |

The annex says the designated Level 3 can delegate v to viii to people holding at least Level 2 in the primary method. It also says each facility must define the nature and frequency of Level 3 involvement, and keep records of it. Software validation sits naturally under ii and viii. A software update is both a procedure question and an equipment question.

## Where could AI touch this?

We could find no mention of AI in the annex. But AI features are arriving in exactly the places it controls: automated defect recognition in radiography and ultrasonic analysis, auto-sizing, and report drafting.

The annex's software line doesn't care what the software does inside, so the same rules apply to an AI feature as to any other. If it does acquisition or analysis, it is validated and version controlled. If it changes with an update, the update is assessed. If it processes an image, the record names it. And the Level 3 activities above are where a firm decides whether AI is allowed to help at all. Authorising a technician to release results is not something to hand to a tool.

If your firm also holds ISO/IEC 17020 inspection accreditation, the new 17020 edition has its own technology clauses. We set them out in [ISO 17020 2026 changes, in plain English](/thinking/iso-17020-2026-changes).

## An example, made up

A ten-person NDT lab runs PAUT on pipeline girth welds. Halfway through a shutdown, the analysis software on one laptop updates itself overnight. The new version changes a default gate setting.

Under the annex, the lab would want to show three things. First, which welds were analysed on which version. Second, that someone with the authority to do so assessed the update before more results were released. Third, how they confirmed sizing was consistent, for example by re-running a reference scan on both versions. A one-page change record, kept with the job, covers all three.

## Is there a 17025:2025?

You may have read that a new edition of ISO/IEC 17025 was published in 2025. As at 13 November 2026, NATA's ISO/IEC 17025 page describes "the current version, ISO/IEC 17025:2017", and we could not find a 2025 edition in NATA's listings. The change NDT labs did get in 2026 is this annex. Check NATA's page and ISO's catalogue for the latest position before acting on anything you read elsewhere.

## What can you do this month?

These are things a careful NDT facility might consider. They are not accreditation advice.

1. **List every piece of software** that acquires, analyses, processes or reports NDT data, with its current version.
2. **Turn off automatic updates** where you can, so updates happen when you decide.
3. **Keep a change record** for each update: old and new version, what could affect detection or sizing, the check you ran, who accepted it, and the date.
4. **Add the record fields** from the table above to your job sheets and report templates.
5. **Write down your CR/DR filter rules.** Which processing functions are allowed, and who can change them.
6. **Check your endorsements** for anyone overseeing PAUT, TOFD or CR/DR on welds.
7. **Write down how your Level 3 is involved**, how often, and what you will keep as a record.

## One question

A question with no pitch attached: who signs off a software update on your PAUT or DR kit today? [warwick@subfrac.com](mailto:warwick@subfrac.com?subject=One%20question)